666-R013 Reference Plate to check photometric accuracy

They have the same dimensions as a microplate with 96 wells and a diameter of 6.6 mm per window (height 14.5 x width 125 x length 85.8 mm). For the five inserted neutral glasses (columns 3-12) the absorption value can be measured per 16 windows. Another 16 windows are without glass (columns 1+2) and are used as a reference.

Please note

The reference plate consists of five neutral galsses with different nominal absorption values. Therefore the linearity of the absorption scale can be checked by plotting the absorption values for each wavelength in a diagram against the measuring values on the DAkkS calibration certificate.

Article-No.666R013
ApplicationReference plate for micro test plate reader to check the photometric accuracy
ContentNeutral glass filter NG11 (0.25 Abs), columns 3+4;
Neutral glass filter NG5 (0.5 Abs), columns 5+6;
Neutral glass filter NG4 (1.0 Abs), columns 7+8;
Neutral glass filter NG3 (1.5 Abs), columns 9+10;
Neutral glass filter NG3 (2.5 Abs), columns 11+12;
Frame without glass (reference filter)
Standard certificationWavelength: 405; 450; 490; 650nm
Slit width: 1nm
with DAkkS calibration certificate
Possible certificationWavelength: all possible between 405 until 890 nm
Over 890 nm possible however, only with Hellma Analytics calibration certificate
Slit width: all possible until 5 nm