667-UV1 Potassium chloride liquid filter to check stray light at 200nm

The potassium chloride liquid filter consists of 12g/l potassium chloride dissolved in pure water. This filter is suited to check the stray light of a spectrometer in the spectral range between 190 nm and 205 nm, its cut off is at about 200 nm.

Please note

Currently there are different methods used in the European Pharmacopoeia (EP) and the United States Pharmacopoeia (USP) to determine the stray light.


  • Testing accoring to EP (European Pharmacopoeia):
Measurement of potassium chloride (UV1) against reference filter pure water (UV12)


  • Testing accoring to USP 857 (U.S. Pharmacopoeia):
Measurement of potassium chloride (UV1, PL 10mm) against reference filter potassium chloride (UV1H, PL 5mm)

Article-no.EP Set: 667100 (667001 (UV1) / 667012 (UV12))
USP- Set: 667100H (667001 (UV1) / 667001H (UV1H))
ApplicationTo check stray light in the spectral range from 190 to 205 nm according to EP and USP
ContentPotassium chloride 12g/l in pure water
Standard certificationCut-off wavelength: 200 nm
Slit width: 2nm
with DAkkS calibration certificate
Possible certificationsCut-off wavelength fixed
Slit width: all until 5nm